Insert the USB-C cable into the charging port on the bottom of the device. Connect to any power supply (max 5V / 2A). Wait until the battery indicator stops scrolling — usually 1–2 hours.
The unit won't run while it's charging. Charge it fully before the first use.
2
Prep the Skin
Clean and completely dry the area where you'll apply pads. Remove any lotion, oil, or sweat — these kill adhesion and reduce conductivity significantly.
Use clean skin with no cuts or rash. Trim hair if needed.
Flat muscle works best. Over bones and joints the pads won't sit flat.
3
Connect Pads (Unit OFF)
Each lead wire = 2 pads. Press snap connectors into both pads firmly. Plug the other end into CH1 or CH2 on the bottom of the device. You can use one or both channels.
Turn the unit off before you plug in or move the wires.
Each wire needs both of its pads on the skin, or it won't run.
4
Apply Pads to Skin
Peel the pad from its protective film by the corner — never the wire. Don't touch the gel with your fingernail. Press firmly on skin, working from center outward to seal all edges.
Keep pads at least 1 inch apart, never overlapping.
Wider apart feels deeper. Closer together feels more focused.
5
Power On
Press the power button on top. The LCD flashes all icons — normal startup. It returns to the last program used. Backlight turns off after 30 seconds of no input.
The unit turns itself off when the timer ends.
6
Select Mode
TENS · EMS · MASS
Press M to cycle through the 3 modes. The active mode indicator lights on the LCD.
TENSPain relief
EMSMuscle stim
MASSRecovery
Switching modes drops both channels back to the lowest intensity.
7
Select Program
P01 → P42
Press ∧ or ∨ to scroll through programs within the current mode. TENS has 12, EMS has 18, MASS has 12 programs.
Not sure which to pick? The Find tab matches programs to how it feels.
8
Set Intensity
CH1: 00 → 20
Press + or − on the CH1 or CH2 side to adjust intensity independently. There are 20 levels.
Start at 1–5 and wait a minute or two before going up. Aim for strong but comfortable.
TENS and Massage feel like tingling. EMS feels like the muscle gently pulling.
9
Set Timer
10 → 60 min
Press the timer button to set session duration (10–60 min, adjustable in increments). Device auto-shuts off when complete.
About 30 minutes per area, once or twice a day.
10
Lock Screen (Optional)
Press briefly (don't hold) to lock. The lock icon [⊐○] appears top-left on LCD. Prevents accidental changes during your session.
To unlock: press briefly again, OR press both − intensity buttons simultaneously while stimulation is active.
11
Finishing Up
When done:
1. Hold 2 seconds to power off. 2. Grab corner of each pad, peel slowly — never pull the wire. 3. Place pads sticky-side on protective film or pad holder. 4. Reseal bag to prevent drying out.
Turn it off before you take the pads off.
Find
What do you need?
Pick how it feels or your goal, or browse all 42 programs.
Frequency map
Where each situation sits
Pulse rate in Hz (hertz, pulses per second). Low rates feel like a deep, slow pulse. High rates feel like a fine tingle.
What does it feel like?
Getting the intensity right, strong but comfortable, matters more than which program you pick.
TENS (transcutaneous electrical nerve stimulation) eases pain. EMS (electrical muscle stimulation) works the muscle. Massage is for recovery.
What's your goal right now?
Acute Pain Relief
Sharp or current pain — sore muscles, injury, post-surgical aching
Chronic / Deep Pain
Ongoing dull ache, long-term soreness, deeper than surface level
Muscle Relaxation
Reduce tension, loosen tight muscles, wind down after a long day
Post-Exercise Recovery
Clear muscle fatigue, flush out soreness after exertion or PT
Muscle Activation / Strengthening
Re-educate or stimulate muscles that are weak or underactive
Warm Up / Circulation
Prime muscles before activity, increase local blood flow
Pad placement
Where the pads go
Choose a body area to see each pad, the channel it belongs to and the target zone.
Select a Body Area
Help
Fixes, safety and care
Quick fixes, the few safety must-knows, and how to keep your pads working.
Must-knows
Pads never go on the head, the front or sides of the neck, or across the chest.
Skip it with a pacemaker or other implanted device, or during pregnancy.
Keep off skin that's broken, irritated or numb.
Turn it off before you put on, move or take off pads.
Near your arm plate, keep pads 5–10 cm (2–4 inches) away, and check with your surgeon or PT (physical therapist) first.
Everything the manual lists
Places: head and face, front or sides of the neck, across the chest, both calves or feet at once, broken, swollen or numb skin, varicose veins, and over any cancerous site.
People: pregnancy, children, and anyone who bleeds internally after an injury or fracture. With heart disease or epilepsy, ask a doctor first. No EMS over a fracture that's still healing.
Times: driving or running machinery, while asleep, in the bath, shower or pool, and during exercise.
Common Problems
Unit on but no pulse from pads
Check intensity — press [+] on CH1 or CH2 and wait 1 minute to feel the pulse.
Only 1 pad connected per wire? Each lead wire needs 2 pads to complete the circuit.
Poor connection? Unplug and firmly replug the wire into the output port. Try a different wire if no change.
Poor pad contact? Pads must be on clean, dry, flat skin — press all edges down firmly.
One pad feels much stronger than the other
Body areas respond differently — try repositioning the weaker-feeling pad slightly.
Wire connection issue — unplug and firmly replug. If the problem persists, test with a different lead wire.
Device cuts off during the session
Low battery — the battery icon will flash before shutdown. Charge the device.
Timer finished — device auto-shuts off when the set time expires. This is normal. Restart and set a new timer.
Device won't power on
Battery too low — charge the device fully before use.
Currently plugged in? You cannot use the device while charging. Unplug the USB-C cable first, then power on.
Can't adjust intensity or change mode
Screen locked — look for the lock icon [⊐○] at the top left of the LCD. Press the power button briefly to unlock, or press both [−] buttons simultaneously during stimulation.
Mode selection issue — press [M] first to select the main mode, then [∧][∨] to scroll programs within it.
Skin turning red or feeling irritated
Intensity too high or session too long — reduce intensity or shorten to 20 minutes. Max recommended is 30 min per area per session.
Same spot every session — rotate pad position slightly each use. If it starts to feel uncomfortable, move the pads.
Pads aren't sticking well
Skin prep — skin must be completely clean and dry. Oil, lotion, and sweat destroy adhesion.
Reconstitution trick (one time only) — if pads are barely sticking, put a small drop of water on a clean fingertip and gently rub it across the entire gel surface. Place gel-side UP on a flat surface and let fully air dry until the water is absorbed and reconstituted. This can only be done once — after that, replace the pads.
Never wipe with tissue or cloth — this strips the gel. Replace pads when gel is melting, deformed, or still not adhering after reconstitution.
How do I store pads so they last longer?
After each use, place the sticky side back on the protective film it came with or the included pad holder. Seal in the resealable bag. Keep away from direct sunlight, heat, and dust — gel picks up dust permanently and loses adhesion.
How TENS Works
TENS delivers small electrical pulses through the skin to nerve fibers, working via two primary mechanisms:
Gate Control Theory: High-frequency stimulation (50–100 Hz) activates large-diameter Aβ sensory fibers, which "close the gate" to pain signals from smaller C and Aδ fibers at the dorsal horn of the spinal cord — providing immediate, local pain relief.
Endorphin Release: Low-frequency stimulation (2–10 Hz) triggers release of endogenous opioids via mu opioid receptors in the brainstem and spinal cord — producing slower-onset but more diffuse and often longer-lasting relief. These mechanisms are distinct: high-frequency TENS is blocked by delta opioid antagonists; low-frequency TENS is blocked by mu opioid antagonists.
EMS targets motor nerves rather than sensory ones, causing involuntary muscle contractions. It bypasses the brain's voluntary control pathway — useful for muscles that are atrophied, weak, or recovering from disuse.
• 35–50 Hz: Near-tetanic (sustained) contraction — strength training effect • 10–25 Hz: Rhythmic contraction/relaxation — endurance and recovery effect • Larger electrode spacing = deeper, more diffuse muscle activation
Source: PMC5683854 (NMES clinical review); PMC8761348 (electrode size and placement)
TENS Near Orthopedic Hardware
Research on titanium and steel orthopedic plates shows TENS produces less than 3°C temperature change at the implant under therapeutic parameters — considered clinically insignificant. TENS uses biphasic alternating current with no net DC, which minimizes charge accumulation at metal surfaces.
• Passive implants (plates, screws) are NOT generally contraindicated for TENS • Place electrodes at least 5–10 cm from the implant site • Stop immediately if burning or unusual sensation near the hardware • Active implants (pacemakers, ICDs) are entirely different — those ARE contraindicated
Studies on TENS in post-operative pain management show:
• Adequate-intensity TENS reduced analgesic consumption by ~26.5% vs placebo (Bjordal 2003, 21 RCTs, n=1,350) • Electrodes placed ~5 cm on either side of the surgical site — not directly over the incision • 20–30 min sessions, multiple times per day, supported by most evidence • Early post-operative application showed better outcomes than delayed start • TENS during and after PT improved range of motion in orthopedic recovery • For rib/thoracic surgery: TENS showed improved respiratory function and lower pain scores vs control
Turn off and disconnect all wires. Wipe with a soft, slightly damp cloth. Never use chemicals (thinner, benzene), and never let water into any port or opening.
Keeping Pads Clean & Lasting Longer
Pad life varies by skin condition, storage, and use frequency. Extend pad life by:
• Always apply to completely clean, dry, hairless skin
• Store sticky-side on protective film or pad holder after each use
• Seal in the resealable bag — prevents drying and dust contamination
• Never expose to sunlight or extreme temperatures
• Don't touch gel with nails or fingertips
• Don't use detergents, soap, or chemicals on the gel surface
Replace pads when: gel is melting, deformed, electrode not sticking flat, or repeatedly losing adhesion.
Pad Reconstitution (One Time Only)
If pads are marginally sticky: put a small drop of water on a clean fingertip and gently rub it across the entire gel surface. Place gel-side UP and let air dry completely until the water is fully absorbed. This can only be done once — after that, replace the pads.
Pad Holder (Included)
Step 1: Remove pads by pulling the corner — never the wire.
Step 2: Stick pads sticky-side down on the holder to protect gel from dust.
Step 3: Wind the lead wires around the holder for organized storage.
Step 4: Store the holder + pads in the case.
Storing the Device
Store in a cool, dry place between 32°F–104°F (0°C–40°C). Keep out of reach of children and pets. Organize lead wires and charging cable with cable ties in the pouch before placing everything back in the case.